AERI Resistance Validation
Equipment Required
- Large phillips screwdriver
- Ball driver set
- Calibrated resistor kit (2 boxes)
Precaution
- The resistor boxes are sensitive to large changes in ambient temperature (e.g. heating from direct sunlight). It is best to conduct the test with the boxes in a sheltered, stable thermal environment. For best results, place the boxes in the location/environment where the test will be conducted about an hour prior to the test.
Procedure
Hardware setup
- Stop data collection and power OFF the instrument:
- Press ctrl-c up to 3 times in the black Cygwin window to terminate ingest.
- Switch off the power bar in the AERI back-end.
- Remove the front panel of the front-end enclosure
- Disconnect the ABB and HBB connectors from the 1.5ft extension cables going to the blackbodies
- Connect the ABB and HBB connectors to the respective ABB and HBB resistor boxes and lay the boxes on the bottom of the enclosure.
- Power ON the instrument
Software setup
- Ensure that instrument is in RS_3 mode. The test will not work in 3body mode
- Ensure that calibKit.connected=false in the file c:\E-AERI\FTSW_EAERI.Config
- Delete the file c:\E-AERI\Data\EAERIHouseKeeping.hk
- Run FTSW_EAERI_GUI.bat
- In the bottom panel, enter
- # of measurements = 1
- # of coadded scans = 9999
- Click Start Acquisition
- Click Stop Acquistion after about 15 minutes
- Verify that the resistance look OK as they are being acquired (see Table 1.)
- 20 records are needed for the analysis. A new record is acquired every ~30 s, so the test should take ~10 mins + initialization
Analysis
- In a cygwin window
- cd /cygdrive/c/E-AERI/Data
- python txt2hkspc.py EAERIHouseKeeping.hk --resistance
- Transfer the created file EAERIHouseKeeping_YYYYMMDD_HHMMSSS.csv to a computer with MS Excel
- Open EAERIHouseKeeping_YYYYMMDD_HHMMSSS.csv in Excel
- Select all
- Copy
- Open AERI_Resistance_Validation_Template.xls
- Paste data into Bring in AERI HK File worksheet
- Save as AERI_Resistance_Check_SNXXX_YYMMDD.xls
- Check to see if the results pass in the Resistance Check Summary worksheet
- If all six thermistor values are IN Spec (green), then the test passes
- If any of the thermistor values are OUT of Spec (red), then the test fails and should be repeated
- Check to see if the 20 records used are stable or drifting (the average is used in the calculation)
- If the values are drifting, then the temperature of the environment of the resistor boxes may not be stable enough
- Repeat the test for a longer time, in a more stable thermal environment if necessary
- If needed, txt2hkspc.py can be modified to output a larger number records than 20
Hardware packup
- Power OFF the instrument
- Disconnect the resistor boxes
- Reconnect the ABB and HBB connectors to each extension cable
- Replace the front panel onto the AERI front-end
- Power ON the instrument and resume data collection
Table 1. Nominal resistance values and acceptable thresholds. Values in Ohms
ABB | HBB | |||||
---|---|---|---|---|---|---|
top | apex | bottom | top | apex | bottom | |
Min R | 98,947 | 124,931 | 799,464 | 2,759.1 | 9,896.0 | 12,494.8 |
Nominal R | 99,000 | 125,000 | 800,000 | 2,760.0 | 9,900.0 | 12,500.0 |
Max R | 99,053 | 125,069 | 800,536 | 2,760.9 | 9,904.0 | 12,505.2 |